Evaluation of the CritiView in a pig model of Abdominal Aortic Occlusion and graded hemorrhage

机译:CritiView在猪腹主动脉阻塞和分级出血模型中的评估

获取原文
获取原文并翻译 | 示例

摘要

We hypothesize that in the presence of reduced oxygen delivery and extraction, blood flow will be redistributed in order to protect the most vital organs (e.g., brain and heart) by increasing their regional blood flow, while O_2 delivery to the less vital organs (e.g., GI tract or urethral wall) will diminish. Evaluation of mitochondrial function in vivo could be done by monitoring the oxidation reduction state of the respiratory chain. Thus, the NADH redox state of less vital organs could serve as an indicator of overall O_2 imbalance as well as an endpoint of resuscitation. We have therefore tested, in a pig model, a new medical device providing real time data on NADH redox state and tissue blood flow- TBF This device contains a modified three way Foley catheter with a fiber optic probe which connects the measurement unit to the tested tissue. Female pigs underwent graded hemorrhage (GH) or Aortic clamping (AC). The main effects of GH started when blood volume decreased by 30%. At 40% blood loss, minimal levels of TBF were correlated to the maximal NADH levels. The values of the 2 parameters returned to baseline after retransfusion of the shed blood. Aortic clamping led to significant decrease in TBF while NADH levels increased. After aortic declamping the parameters recovered to normal values. Due to the short length of the urethra in female pigs and the instable contact between the probe and the tissue, inconsistency of the responses was observed. Our preliminary results show that the CritiView may be a useful tool for the detection of body O_2 imbalance.
机译:我们假设在氧气输送和抽取减少的情况下,血流将被重新分配,以通过增加局部血液流量来保护最重要的器官(例如,大脑和心脏),而将O_2输送到次要的器官(例如, ,胃肠道或尿道壁)将减少。体内线粒体功能的评估可通过监测呼吸链的氧化还原状态来完成。因此,次要器官的NADH氧化还原状态可以作为总体O_2失衡的指标,也可以作为复苏的终点。因此,我们在猪模型中测试了一种新的医疗设备,该设备可提供有关NADH氧化还原状态和组织血流量的实时数据-TBF该设备包含带有光纤探头的改进的三向Foley导管,该探头将测量单元连接到被测组织。雌性猪经历了分级出血(GH)或主动脉夹闭(AC)。 GH的主要作用始于血容量减少30%。失血率为40%时,最小的TBF水平与最大的NADH水平相关。再流血后,这两个参数的值返回到基线。主动脉夹闭导致TBF显着下降,而NADH水平升高。主动脉放松后,参数恢复到正常值。由于雌性猪的尿道长度短以及探针和组织之间的接触不稳定,因此观察到反应不一致。我们的初步结果表明,CritiView可能是检测体内O_2不平衡的有用工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号