There is a pressing need for a passive optical fiber dosimeter probe for use in real-time monitoring of radiation dose delivered to clinical radiation therapy patients. An optical fiber probe using radiochromic material has been designed and fabricated based on a thin film of the radiochromic material on a dielectric mirror. Measurements of the net optical density vs. time before, during, and after irradiation at a rate of 500cGy/minute to a total dose of 5 Gy were performed. Net optical densities increased from 0.2 to 2.0 for radiochromic thin film thicknesses of 2 to 20 urn, respectively.
机译:迫切需要一种无源光纤剂量计探头,用于实时监测交付给临床放射治疗患者的放射剂量。已经基于介电镜上的辐射致变色材料的薄膜设计和制造了使用辐射致变色材料的光纤探针。以500cGy /分钟的速率照射到总剂量5 Gy的辐射之前,之中和之后的净光密度与时间的关系进行了测量。对于2至20微米的放射致变色薄膜,净光学密度从0.2增至2.0。