首页> 外文会议>Optical coherence tomography and coherence domain optical methods in biomedicine XV >In-vivo Human retina imaging with 5um axial resolution, at 92000 A-scans/s with liim Spectral Domain OCT system
【24h】

In-vivo Human retina imaging with 5um axial resolution, at 92000 A-scans/s with liim Spectral Domain OCT system

机译:使用liim光谱域OCT系统以92000 A-scan / s的速度进行5um轴向分辨率的体内人类视网膜成像

获取原文
获取原文并翻译 | 示例

摘要

We have outfitted a 1060nm Spectral Domain Optical Coherence Tomography system with a prototype, high speed infrared linear array camera and a custom spectrally reshaped superluminescent diode to achieve 5 μm axial resolution at 91,911 A-scans/s image acquisition rate in-vivo in the human retina. 4dB loss of sensitivity was observed as a result of the reduced integration time (7μs) of the fast camera as compared to similar commercially available cameras with 14μs integration time and 47kHz readout rate. Fewer motion artefacts were observed in the retinal images acquired with the fast camera, while the higher axial resolution along with deeper penetration allowed for improved visualization of fine morphological details such as retinal and choroidal capillaries and the deep choroidal structure.
机译:我们为1060nm光谱域光学相干断层扫描系统配备了原型,高速红外线性阵列相机和定制的光谱重塑超发光二极管,以实现91,911 A-scan / s的人体体内图像采集速率的5μm轴向分辨率视网膜。与具有14μs积分时间和47kHz读出速率的类似商用相机相比,由于快速相机的积分时间(7μs)减少,导致灵敏度降低了4dB。用快速相机采集的视网膜图像中观察到的运动伪影较少,而较高的轴向分辨率以及更深的穿透力使精细的形态学细节(如视网膜和脉络膜毛细血管以及深脉络膜结构)的可视化得以改善。