A piezoelectric detector with cylindrical shape for photoacoustic section imaging is characterized. This detector is larger than the imaging object in direction of the cylinder axis, giving rise to its integrating properties. Its focal volume has the shape of a slice and the acquisition of signals for one section image requires rotation of an object about an axis perpendicular to this slice. Image reconstruction from the signals requires the application of the inverse Radon transform. It is shown that implementing the Abel transform is a suitable step in data processing, allowing speeding up the data acquisition since the scanning angle can be reduced. The resolution of the detector was estimated in directions perpendicular and parallel to the detection plane. An upper limit for the out of plane resolution is given and section images of a zebra fish are shown.
机译:表征了用于光声截面成像的圆柱形压电检测器。该检测器在圆柱轴方向上大于成像对象,从而产生其积分特性。其焦点体积具有切片的形状,并且获取一个截面图像的信号需要使对象绕垂直于该切片的轴旋转。从信号重建图像需要应用Radon逆变换。结果表明,实现Abel变换是数据处理中的合适步骤,由于可以减小扫描角度,因此可以加快数据获取速度。在垂直于和平行于检测平面的方向上估计检测器的分辨率。给出了平面外分辨率的上限,并显示了斑马鱼的剖面图像。