Migration of endovascular stent grafts for abdominal aortic aneurysm (AAA) is a serious complication that can result in aneurysm sac repressurisation and rupture. Secure fixation of stent-grafts in the proximal neck and the iliac legs is crucial to prevent this complication. We outline an in vitro method of investigating the resistance to migration of both a commercially available (AneuRx, Medtronic, Santa Rosa, Ca, USA) and homemade device. The devices were implanted in a silicone rubber mock AAA model and subjected to physiological pulsatile flow. The fixation strength of the AneuRx device was 8.52±0.5N and the fixation force for the homemade stent-graft was 4.62+0.25N. These results highlight the importance of longitudinal strength in devices without hooks or barbs on the proximal stent.
机译:腹血管支架移植物用于腹主动脉瘤(AAA)的迁移是一种严重的并发症,可能导致动脉瘤囊反应和破裂。在近端颈部和髂骨中的支架移植物固定固定对于防止这种并发症是至关重要的。我们概述了一种调查抗迁移的体外方法,抵抗市售(Aneurx,Medtronic,Santa Rosa,CA,USA)和自制装置。将该装置植入硅橡胶模型AAA模型中并进行生理脉动流。 ANEURX器件的固定强度为8.52±0.5N,自制支架移植物的固定力为4.62±0.25N。这些结果突出了近端支架上没有钩子或倒钩的装置中纵向强度的重要性。