首页> 外文会议>ASME summer bioengineering conference;SBC2012 >MEASUREMENT OF FLOW THROUGH HYPODERMIC NEEDLES WITH MODIFIED ENTRANCES FOR REDUCED HEMOLYSIS
【24h】

MEASUREMENT OF FLOW THROUGH HYPODERMIC NEEDLES WITH MODIFIED ENTRANCES FOR REDUCED HEMOLYSIS

机译:降低渗透性的经过修饰的通透性皮下注射针的测量

获取原文

摘要

These experimentally determined flow velocities in needles with modified entrance geometries validate the trends found previously in CFD simulations, confirming that entrance losses in hypodermic needles represent a significant portion of the overall losses through the needles. Because the modified entrance geometries affect both the mean velocity through the needle and the distribution of stresses near the entrance, these conditions present challenging benchmark cases suitable for testing hemolysis prediction models.
机译:这些通过实验确定的入口几何形状已改变的针头流速,证实了先前在CFD模拟中发现的趋势,从而证实皮下注射针头的入口损失占整个针头总损失的很大一部分。由于修改后的入口几何形状会影响通过针头的平均速度以及入口附近应力的分布,因此这些条件提出了具有挑战性的基准案例,适合测试溶血预测模型。