首页> 外文会议>2011 International Conference on Electric Technology and Civil Engineering >Statistical analysis of a HPLC method for microcystins determination in drinking water
【24h】

Statistical analysis of a HPLC method for microcystins determination in drinking water

机译:高效液相色谱法测定饮用水中微囊藻毒素的统计分析

获取原文

摘要

Statistical analysis of high performance liquid chromatography (HPLC) method for the determination of 0.1∼2.0mg·L−1 microcystins (MCs) in drinking water was investigated to optimize the instrument operating conditions and reduce the detection limit. The detection limits were obtained by two ways. One through method detection limit (MDL) and the other through detection limit from the calibration curve (DTC). The Cochran test results revealed the homoscedasticity of the HPLC method. The effect of system drift on the calibration curve was noticed by the lack of fit (LOF) test. MDL yielded a less reliable detection limit than DTC. The detection limit of HPLC method for MC-LR was 0.009mg·L−1 under recommended operating conditions.
机译:对高效液相色谱法测定饮用水中0.1〜2.0mg·L -1 微囊藻毒素(MCs)的统计分析进行了研究,以优化仪器操作条件并降低检测限。检测限是通过两种方法获得的。一种通过方法的检测极限(MDL),另一种通过校准曲线的检测极限(DTC)。 Cochran测试结果显示了HPLC方法的均方差性。缺少拟合(LOF)测试注意到了系统漂移对校准曲线的影响。 MDL产生的检测极限不如DTC可靠。在推荐的操作条件下,MC-LR的HPLC方法检出限为0.009mg·L -1

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号