Power Line Communication (PLC) is a communication system for transmitting data through power lines instead of using conventional local area network cables. However, signals from a PLC modem superimposed on the alternating current electric power could generate noise to the power system of medical devices and thus induce malfunction. To investigate the effects of PLC use on electronic medical devices, we used six types of PLC modems and 14 kinds of medical devices. The voltages and distribution ranges of the noise superimposed on the electric power voltage by use of PLC and the intensity of the electromagnetic field radiated from each PLC modem were measured. The maximum noise voltage superimposed was 17.68 V, and it appeared every 5.5 msec. The frequency distribution ranged from 2 to 30 MHz for almost all subject modems, but one type modem exceeded 30MHz. The electric field intensity radiated from each PLC modem was less than 1 V/m at 1 m. No interference by the electric field radiated from a PLC modem was observed on any of the subject medical devices, although artifact noise was observed on two pieces of ultrasonic diagnostic equipment when their power was supplied from an outlet between the master and terminal PLC modems and when the operator held the probe. This type of malfunction can be eliminated by dividing power lines for a medical device from the ones for PLC.
机译:电力线通信(PLC)是用于通过电力线传输数据而不是使用传统的局域网电缆的通信系统。然而,来自PLC调制解调器的信号叠加在交流电电力上可能产生噪声对医疗设备的电力系统,从而产生诱导故障。为了调查PLC对电子医疗器械的影响,我们使用了六种类型的PLC调制解调器和14种医疗设备。测量通过使用PLC叠加在电力电压上的噪声的电压和分配范围和从每个PLC调制解调器辐射的电磁场的强度。叠加的最大噪声电压为17.68 V,并显示每5.5毫秒。几乎所有主题调制解调器的频率分布范围为2到30 MHz,但是一个类型的调制解调器超过30MHz。从每个PLC调制解调器辐射的电场强度在1米处小于1V / m。在任何受试者医疗器械上观察到从PLC调制解调器辐射的电场的干扰,尽管在从主电源和终端PLC调制解调器之间的出口供应时,在两件超声诊断设备上观察到伪影噪声。操作员持有探针。通过从PLC的电源分开用于医疗设备的电力线可以消除这种情况。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号