首页> 外文会议>Medical Imaging 1999: PACS Design and Evaluation: Engineering and Clinical Issues >Frequency and failure of high-resolution monitors in a filmless imaging department
【24h】

Frequency and failure of high-resolution monitors in a filmless imaging department

机译:无胶卷成像部门中高分辨率监视器的频率和故障

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The purpose of this study was to assess the rate of failure of the high resolution monitors used for primary diagnosis in a filmless radiology department and to analyze the type of problems encountered as well as the action taken to report the monitors. Data were collected from PACS service logs to determine rates of monitor adjustment and replacement, the symptoms reported, and the action taken. Analysis of monitor service records demonstrated a high rate of monitor replacement resulting in a relatively short 'life expectancy of 2.4 years. The result of this study support our subjective impression and those of other colleagues in the PACS community of an unacceptably high monitor failure rate with 2,000 pixel monitors used for primary diagnosis. Requests for proposals for PACS and service contracts must carefully specify requirements for monitor images quality and conditions under which the vendor is required to replace these monitors. !9
机译:摘要:这项研究的目的是评估用于无胶片放射科的主要诊断的高分辨率监视器的故障率,并分析遇到的问题的类型以及报告监视器所采取的措施。从PACS服务日志中收集数据,以确定监视器调整和更换的速率,报告的症状以及采取的措施。监控器服务记录的分析表明,监控器更换率很高,导致“预期寿命”相对较短,为2.4年。这项研究的结果支持了我们和PACS社区其他同事的主观印象,即使用2,000像素监视器进行初步诊断的监视器故障率极高。要求PACS和服务合同的提案的请求必须仔细指定监视器图像质量的要求以及要求供应商更换这些监视器的条件。 !9