Low-frame-rate pulsed fluoroscopy vs low-exposure-rate continuous fluoroscopy: a comparison

机译:低帧率脉冲荧光透视与低曝光率连续荧光透视:比较

获取原文
获取原文并翻译 | 示例

摘要

Abstract: Although improved video camera sensitivity has been responsible for reducing the radiation exposure rate in conventional fluoroscopy substantially over the last two decades, considerable further improvement is possible. Over the same period two methods for reducing fluoroscopic exposure have been implemented by various researchers: pulsed X-ray irradiation and progressive video acquisition of the image intensifier output at low frame rate with exposure per frame comparable to conventional fluoroscopy; and continuous irradiation with a combination of low exposure rate, high detector gain and X-ray beam filtration. Neither has come into widespread clinical use. This paper proposes a model of fluoroscopic patient exposure and compares pulsed fluoroscopy with other methods of exposure reduction. The implications of this model for fluoroscopic exposure rate optimization are discussed.!9
机译:摘要:尽管在过去的二十年中,提高摄像机的感光度已经可以降低传统荧光检查的辐射曝光率,但仍有可能取得更大的进步。在同一时期,各种研究人员已经实施了两种减少荧光透视曝光量的方法:脉冲X射线辐照和渐进式视频采集图像增强器输出的低帧频,每帧曝光量可与常规荧光透视相媲美;低曝光率,高检测器增益和X射线束过滤相结合的连续照射。两者均未广泛用于临床。本文提出了一种荧光透视患者暴露模型,并将脉冲荧光透视与其他减少暴露的方法进行了比较。讨论了该模型对荧光镜曝光率优化的意义。9