首页> 外文会议>Medical Imaging 1993: Physics of Medical Imaging >Evaluation of x-ray sources for mammography
【24h】

Evaluation of x-ray sources for mammography

机译:乳腺摄影的X射线源评估

获取原文
获取原文并翻译 | 示例

摘要

Abstract: A computational approach is being developed for the evaluation of mammographic imaging system performance. This approach takes into account both the spatial frequency properties and the x-ray spectral characteristics of the system being evaluated. The initial version of the program that implements the approach has been used to evaluate a conventional mammography source assembly for several breast thicknesses, and to compare the conventional tube and filter combination to alternatives that have been suggested for the imaging of breasts that are thicker or more dense than average. It has also been used to study the effect of varying the thickness of the molybdenum filter in the conventional system. The parameters calculated include contrast, average glandular dose, tube load, and a figure of merit, SNR$+2$//Dose. The calculations confirm the strong dependence of system performance on both tube potential and breast thickness for the standard system, and indicate that alternative designs can improve performance in the imaging of thicker or more dense breasts. The study of filter thickness shows that, of the four parameters calculated, only tube load is strongly affected by filter thickness.!14
机译:摘要:目前正在开发一种计算方法,用于评估乳腺X线摄影成像系统的性能。该方法同时考虑了被评估系统的空间频率特性和X射线光谱特性。实施该方法的程序的初始版本已用于评估几种乳房厚度的常规乳房X线照相放射源组件,并将常规的导管和过滤器组合与建议用于成像更厚或更厚乳房的替代方案进行比较比平均密度高。它也已被用来研究在传统系统中改变钼过滤器厚度的效果。计算的参数包括对比度,平均腺体剂量,管负荷和品质因数SNR $ + 2 $ // Dose。这些计算确认了标准系统的系统性能对电子管电势和乳房厚度的强烈依赖性,并表明替代设计可以改善较厚或更密实乳房的成像性能。对过滤器厚度的研究表明,在计算出的四个参数中,只有管负载受过滤器厚度的影响很大!14

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号