首页> 外文会议>Medical Imaging 1993: Physics of Medical Imaging >Image intensifier-based volume tomographic angiography imaging system: work in progress
【24h】

Image intensifier-based volume tomographic angiography imaging system: work in progress

机译:基于图像增强器的体层血管造影成像系统:正在进行中

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The construction of the imaging system includes: (1) replacing the linear detector of the CT scanner with an image intensifier coupled to a TV camera, (2) modifying the gantry to mount the image intensifier, (3) replacing the system computer by a PC, (4) designing and constructing the interfacing and data acquisition system and (5) developing software for system control and data acquisition. The system consists of a x-ray tube and an image intensifier that are separately mounted on a gantry. The tube- detector set can be rotated over 360$DGR in the gantry. Two sets of images acquired with this imaging system indicate that two types of distortion occur in image intensifiers, S distortion and pincushion distortion. In the future, these two types of distortions need to be corrected prior to three-dimensional (3-D) reconstruction.!18
机译:摘要:成像系统的结构包括:(1)用与电视摄像机相连的图像增强器替换CT扫描仪的线性检测器;(2)修改机架以安装图像增强器;(3)替换系统计算机通过PC,(4)设计和构建接口和数据采集系统,以及(5)开发用于系统控制和数据采集的软件。该系统由分别安装在机架上的X射线管和图像增强器组成。可以将管检测器套件在机架中旋转超过360 $ DGR。用该成像系统采集的两组图像表明,图像增强器中发生两种类型的失真:S失真和枕形失真。将来,在进行三维(3-D)重建之前,需要校正这两种类型的失真!18