Ionic mobility dispersion in KCD digital radiography: dopants and physical parameters

机译:KCD数字射线照相术中的离子迁移率色散:掺杂剂和物理参数

获取原文
获取原文并翻译 | 示例

摘要

Abstract: The spatial resolution of the Kinestatic Charge Detector (KCD) for digital radiography is limited by mobility dispersion when the detector operates with noble gases such as xenon or krypton. The magnitude and dependence on drift distance of the peak widths of ionic signal pulses produced in the KCD provide a measure of mobility dispersion. These parameters have been measured in a KCD filled with krypton gas at a pressure of 60 atm, and in the same gas alternately doped with 1.7% of each of the following amine additives: ammonia, methylamine, dimethylamine and trimethylamine. The ionization potentials of these dopants are 10.2, 9.0, 8.2 and 7.8 ev, respectively. While the undoped medium exhibited significant mobility dispersion, all four of the amine- doped media showed dramatic reduction or elimination of mobility dispersion.!19
机译:摘要:当检测器与稀有气体(例如氙或k)一起工作时,用于数字射线照相的运动静电荷检测器(KCD)的空间分辨率受到迁移率色散的限制。 KCD中产生的离子信号脉冲的峰值宽度的大小和对漂移距离的依赖性提供了迁移率色散的度量。这些参数是在充满CD气,压力为60 atm的KCD中测量的,并且在同一气体中交替掺杂了1.7%的以下每种胺添加剂:氨,甲胺,二甲胺和三甲胺。这些掺杂剂的电离电势分别为10.2、9.0、8.2和7.8ev。虽然未掺杂的介质显示出显着的迁移率分散性,但所有四种胺掺杂介质均显示出显着的迁移率分散性降低或消除。