首页> 外文会议>Measurement of toxic and related air pollutants >Accuracy considerations of portable electrochemeical NOX analyzers
【24h】

Accuracy considerations of portable electrochemeical NOX analyzers

机译:便携式电化学NOX分析仪的精度考量

获取原文
获取原文并翻译 | 示例

摘要

Two key components contributing to measurement errors of electrochemical analyzers are discussed. These are the sample conditiong system and the elelctrochemeical nitric oxide and nitrogen dioxide sensors. The problems associated with various types of conditioning systems are discussed and some experimental results are presented using analyte spiking methods. Permeation drier based systems are shown to cause the smallest loss of the analyte. Two major problems of the NO and NO2 sensors are examined. The first problem deals with the significant effect of temperature on the sensor and its associated interference rejection filter. The requirement for maintaining sensor and filter temperature below 30 deg. C is demonstrated. The second deals with the saturation and drift considerations caused by over exposure to the gas. The significance of capillar size to minimize drifit for difusion sensors is discussed. Experimental results are presented and discussed with a view to the recently published EPA CTM-022 Method.
机译:讨论了导致电化学分析仪测量误差的两个关键因素。这些是样品处理系统以及电化学一氧化氮和二氧化氮传感器。讨论了与各种类型的调节系统相关的问题,并使用分析物加标方法介绍了一些实验结果。基于渗透干燥器的系统显示出导致分析物损失最小。研究了NO和NO2传感器的两个主要问题。第一个问题涉及温度对传感器及其相关干扰抑制滤波器的重大影响。保持传感器和过滤器温度低于30度的要求。演示了C。第二部分处理了由于过度暴露于气体而引起的饱和度和漂移问题。讨论了毛细管尺寸对最小化扩散传感器漂移的重要性。针对最近发布的EPA CTM-022方法,介绍和讨论了实验结果。