A comparison feld study of protable gas chromatographs (PGC) was conducted during which foru commercially-available PGCs were operated downwind of a source generator. Concurrent colocated PGC and canister samples were also collected and analyzed later by Method TO-14. Each instrument also analyzed audit standards of TO-14 target compountds at five different concentrations in the low ppb range. These were furnished in proessurized canisters. Adverse weather complicated the study. Samples were collected, when possible, during periods of stable wind veelocity and minimal proecipitation. Agreement between PGs and the reference method was imperfect for plume analysis data bacause of weather conditions. Better perforamance was observed when audit standards were analzed.
机译:进行了可比较气相色谱仪(PGC)的对比研究,在此过程中,市场上可买到的PGC在源发生器的顺风方向运行。同时收集了并置的PGC和罐样品,并随后通过方法TO-14进行了分析。每个仪器还分析了低ppb范围内五个不同浓度下的TO-14目标化合物的审核标准。这些都装在经过鉴定的罐中。不利的天气使这项研究复杂化。在可能的情况下,在稳定的风速和最少的降水期间收集样品。 PG和参考方法之间的一致性因天气状况而对羽流分析数据不完善。分析审计标准时,观察到更好的性能。