首页> 外文会议>Measurement of toxic and related air pollutants >The application of high speed gas chromatography to air analysis
【24h】

The application of high speed gas chromatography to air analysis

机译:高速气相色谱法在空气分析中的应用

获取原文
获取原文并翻译 | 示例

摘要

The demand for frequent sampling (2 to 5 minutes) approaching a continuous mode has generated interest in high speed GC analysis. This paper describes an apparatus which produces a minimum injection bandwidth (10 to 50 msec) which provides the primary means of attaining good column efficiency while achieving High Speed GC Analysis. THis FastGC inlet which is mounted on a Varian 3400 Gas Chromatiograph is described and its operation is detalied from the cryofocussing of the dample onto a short Nickel tube to the 100,000 deg C /sec desorption to the 8 Meter x 0.32 mm capillary coumn. The applications described in the paper include the analysis of chlorinated hydrocarbons, polar organics, an ozone precursor standard (C2 to C13), and BTEX's Data regarding the recovery, response linearity, and retention time and area count precision are also reported.
机译:接近连续模式的频繁采样(2至5分钟)的需求引起了对高速GC分析的兴趣。本文介绍了一种产生最小进样带宽(10到50毫秒)的设备,该设备提供了在实现高速GC分析的同时获得良好色谱柱效率的主要手段。描述了安装在Varian 3400气相色谱仪上的FastGC进样口,其操作从将冷气聚焦在短镍管上的阻尼器低温聚焦到100,000摄氏度/秒的解吸到8米x 0.32毫米毛细管柱上都是不可行的。该论文中描述的应用包括氯代烃,极性有机物,臭氧前体标准物(C2至C13)的分析,还报道了BTEX关于回收率,响应线性,保留时间和面积计数精度的数据。