INVESTIGATION ON BIOMEDICAL MEASUREMENT ACCURACY IN ELECTRODE-SKIN TESTS

机译:电极皮肤测试中生物医学测量准确性的调查

获取原文
获取原文并翻译 | 示例

摘要

The paper investigates three sources of potential error in measurements that may be taken as part of a clinical protocol using Electrical Impedance Tomography (EIT). The findings of an investigation show that the error of two types of electrode used (Stainless Steel and ECG) could produce 13.5-44.5%, 1.5-34.8% and 9.9-83.7% errors in repeatability, contact drift and electrode mismatching studies, respectively. Such widely varying errors challenge EIT imaging both when measurements are taken at different clinical visits and even at the same visit. The paper discusses the consequence of such measurement inaccuracies and proposes a "best practice" showing how to reduce these errors.