A Reproduction of Inflow Restriction in the Mock Circulatory System to Evaluate a Hydrodynamic Performance of a Ventricular Assist Device in Practical Conditions

机译:模拟循环系统中的流入限制的再现,以评估实际条件下的心室辅助装置的水动力性能

获取原文
获取原文并翻译 | 示例

摘要

A novel in vitro mock circulatory system, which enables to reproduce an inflow restriction, simulating blood volume pooling due to heart failure, was developed to evaluate a hydrodynamic performance of a pulsatile ventricular assist device (VAD) in practical conditions. The concept of this development was motivated by a difference of an inflow restriction between in vitro and in vivo environments. The major idea of this study is to reproduce an inflow restriction by using a centrifugal pump placed at an inflow side of a left ventricular model instead of a constant head reservoir in a conventional circuit. In the novel circuit, the maximum flow rate was obtained at lower systolic fraction as compared with a conventional circuit. This similar tendency by the novel one was observed in an acute animal experiment in sheep. This result suggests that a new mock circuit is effective to confirm a practical drive strategy of the VAD for various diseased conditions.
机译:开发了一种新颖的体外模拟循环系统,该系统能够重现入流限制,模拟由于心力衰竭引起的血容量积,从而在实际条件下评估脉动性心室辅助装置(VAD)的流体力学性能。这种发展的概念是由于体外和体内环境之间的流入限制不同而引起的。这项研究的主要思想是通过使用放置在左心室模型流入侧的离心泵代替常规回路中的恒定压头水库来再现流入限制。在新颖的回路中,与常规回路相比,在较低的收缩分数下获得了最大流速。在绵羊的急性动物实验中观察到了这种新趋势。该结果表明,新的模拟电路可以有效地确定各种疾病条件下VAD的实用驱动策略。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号