首页> 外文会议>Design and quality for biomedical technologies X >Traceable working standards with SI units of radiance for characterizing the measurement performance of investigational clinical NIRF imaging devices
【24h】

Traceable working standards with SI units of radiance for characterizing the measurement performance of investigational clinical NIRF imaging devices

机译:具有SI辐射单位的可追溯工作标准,用于表征临床用NIRF成像设备的测量性能

获取原文
获取原文并翻译 | 示例

摘要

All medical devices for Food and Drug market approval require specifications of performance based upon Internationa] System of Units (SI) or units derived from SI for reasons of traceability. Recently, near-infrared fluorescence (NIRF) imaging devices of a variety of designs have emerged on the market and in investigational clinical studies. Yet the design of devices used in the clinical studies vary widely, suggesting variable device performance. Device performance depends upon optimal excitation of NIRF imaging agents, rejection of backscattered excitation and ambient light, and selective collection of fluorescence emanating from the fluorophore. There remains no traceable working standards with SI units of radiance to enable prediction that a given molecular imaging agent can be detected in humans by a given NIRF imaging device. Furthermore, as technologies evolve and as NIRF imaging device components change, there remains no standardized means to track device improvements over time and establish clinical performance without involving clinical trials, often costly. In this study, we deployed a methodology to calibrate luminescent radiance of a stable, solid phantom in SI units of mW/cm~2/sr for characterizing the measurement performance of ICCD and IsCMOS camera based NIRF imaging devices, such as signal-to-noise ratio (SNR) and contrast. The methodology allowed determination of superior SNR of the ICCD over the IsCMOS system; comparable contrast of ICCD and IsCMOS depending upon binning strategies.
机译:所有出于食品和药品市场批准的医疗设备都需要基于国际单位制(SI)或源自SI的单位的性能规格,以进行追溯。近来,各种设计的近红外荧光(NIRF)成像设备已经出现在市场和临床研究中。然而,临床研究中使用的设备设计差异很大,表明设备性能各不相同。设备性能取决于NIRF成像剂的最佳激发,反向散射激发和环境光的抑制以及荧光团发出的荧光的选择性收集。尚无具有SI辐射度单位的可追溯工作标准,无法预测给定的NIRF成像设备可在人体中检测到给定的分子成像剂。此外,随着技术的发展以及NIRF成像设备组件的变化,没有标准化的方法可以随时间跟踪设备的改进并建立临床性能,而无需进行临床试验,而这往往是昂贵的。在这项研究中,我们采用了一种以mW / cm〜2 / sr的SI单位校准稳定的实心模型的发光辐射的方法,以表征基于ICCD和IsCMOS相机的NIRF成像设备的测量性能,例如信噪比。噪声比(SNR)和对比度。该方法可以确定ICCD优于IsCMOS系统的SNR。 ICCD和IsCMOS的可比性取决于分箱策略。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号