Validation of MTF measurement for CBCT system using Monte Carlo simulations

机译:使用Monte Carlo模拟验证CBCT系统的MTF测量

获取原文
获取原文并翻译 | 示例

摘要

To evaluate the spatial resolution performance of cone beam computed tomography (CBCT) system, accurate measurement of the modulation transfer function (MTF) is required. This accuracy depends on the MTF measurement method and CBCT reconstruction algorithms. In this work, the accuracy of MTF measurement of CBCT system using wire phantom is validated by Monte Carlo simulation. A Monte Carlo simulation software tool BEAMnrc/EGSnrc was employed to model X-ray radiation beams and transport. Tungsten wires were simulated with different diameters and radial distances from the axis of rotation. We adopted filtered back projection technique to reconstruct images from 360° acquisition. The MTFs for four reconstruction kernels were measured from corresponding reconstructed wire images, while the ram-lak kernel increased the MTF relative to the cosine, hamming and hann kernel. The results demonstrated that the MTF degraded radially from the axis of rotation. This study suggested that an increase in the MTF for the CBCT system is possible by optimizing scanning settings and reconstruction parameters.
机译:若要评估锥束计算机断层扫描(CBCT)系统的空间分辨率性能,需要对调制传递函数(MTF)进行精确测量。此精度取决于MTF测量方法和CBCT重建算法。在这项工作中,通过蒙特卡洛仿真验证了使用线体模的CBCT系统MTF测量的准确性。使用Monte Carlo模拟软件工具BEAMnrc / EGSnrc对X射线辐射束和传输进行建模。模拟了不同直径和距旋转轴的径向距离的钨丝。我们采用滤波反投影技术从360°采集中重建图像。从相应的重构导线图像中测量了四个重构内核的MTF,而Ram-lak内核相对于余弦,汉明和汉恩内核增加了MTF。结果表明,MTF从旋转轴径向退化。这项研究表明,通过优化扫描设置和重建参数,可以增加CBCT系统的MTF。

著录项

  • 来源
  • 会议地点 San Francisco CA(US)
  • 作者单位

    School of Precision Instrument and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China;

    School of Precision Instrument and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China,Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, Tianjin 300072, China;

    School of Precision Instrument and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China,Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, Tianjin 300072, China;

    School of Precision Instrument and Optoelectronics Engineering, Tianjin University, Tianjin 300072, China,Tianjin Key Laboratory of Biomedical Detecting Techniques and Instruments, Tianjin 300072, China;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Cone beam computed tomography; modulation transfer function; Monte Carlo simulation; BEAMnrc; EGSnrc;

    机译:锥形束计算机断层扫描;调制传递函数;蒙特卡罗模拟; BEAMnrc; EGSnrc;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号