In this study, we present comparison of human and software reading of contrast-detail mammography (CDMAM) phantom images for a digital mammography system. The human readings were conducted through CDMAM images using 3 Megapixel and 5 Megapixel medical monitors. The phantom was positioned between Poly methyl methacrylate (PMMA) blocks of 20mm thickness. A set of 8 digital radiographs were taken using the automated exposure control (AEC-Opdose). The phantom was slightly repositioned between each exposure, always with the smaller details close to the chest wall side. Automated readings were carried out with CDMAM 3.4 Analyser software V.2.8. (CDCOM) and human readings were evaluated by three observers. As a result of this study, CDMAM 3.4 Analyser software is able to provide results significantly better than those estimated by human observers. Additionally, human readings were better for 5 Megapixel medical monitor. We can note that quality control tests of mammography systems and clinical evaluation of mammogram should be evaluated with 5 Megapixel medical monitor.
机译:在这项研究中,我们比较了数字乳腺X线摄影系统的对比细节乳腺X线摄影(CDMAM)体模图像的人机和软件读取情况。使用3百万像素和5百万像素医疗监视器通过CDMAM图像进行人类读数。幻影位于厚度为20mm的聚甲基丙烯酸甲酯(PMMA)块之间。使用自动曝光控制(AEC-Opdose)拍摄了8张数字X射线照片。每次曝光之间,幻影都要稍稍重新定位,并且始终在靠近胸壁一侧的地方放置较小的细节。使用CDMAM 3.4分析仪软件V.2.8进行自动读数。 (CDCOM)和人类读数由三位观察员评估。这项研究的结果是,CDMAM 3.4分析仪软件能够提供比人类观察者估计的结果明显更好的结果。此外,对于5百万像素医疗显示器,人类读数更好。我们可以注意到,应该使用5兆像素医疗监视器来评估乳房X线照相系统的质量控制测试和乳房X线照片的临床评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号